Testing errors in measuring the effective orifice diameter as a new sizing parameter of surgical valve prostheses
Letter to the Editor

Testing errors in measuring the effective orifice diameter as a new sizing parameter of surgical valve prostheses

Astrid Gerritje Maria van Boxtel ORCID logo, Tjark Ebels ORCID logo

University Medical Centre Groningen, Groningen, The Netherlands

Correspondence to: Prof. Tjark Ebels, MD, PhD. University Medical Centre Groningen, Hanzeplein 1, 9732GZ Groningen, The Netherlands. Email: tjarkebels@mac.com.

Comment on: Wu C, Green C, Marquez S, et al. Effective orifice diameter: a new sizing parameter of surgical valve prostheses to inform valve selection. Ann Cardiothorac Surg 2024;13:283-90.



Submitted Feb 03, 2026. Accepted for publication Aug 31, 2026. Published online Sep 29, 2026.

doi: 10.21037/acs-2026-1-0038


We are very thankful to Wu and the rest of the International Organization for Standardization (ISO) Cardiac Valves Working Group (the Working Group) for their efforts to increase transparency and better inform surgeons about the orifice sizes of prosthetic heart valves with their 2024 paper entitled “Effective orifice diameter: a new sizing parameter of surgical valve prostheses to inform valve selection” (1). However, the paper contains essential errors in the chosen test methodology. These errors cause published effective orifice area’s (EOA’s) to be too small in the smaller sizes of the tested prosthetic valves in this paper.

The purpose of the study was to determine whether the effective orifice diameter (ID-effective) could be derived from the EOA measured under steady flow conditions in a repeatable and reproducible manner as described in the ISO Publicly Available Specification (PAS) 7020:2023 (2). In their paper (1), Wu et al. concluded that this test method provided a repeatable and reproducible characterization and the results of the steady flow round-robin study were reported, naming the manufacturers, brands and sizes.

However, we noticed in the test set-up description, that an outlet tube of 35 mm internal diameter was used for all sizes of prosthetic valves, which is in violence of the stipulations of the ISO PAS (2). The PAS describes, that for valves with an inflow orifice diameter of 20.0 mm or less a 25 mm outlet tube should be used. Using the same sized outlet tube for all sizes of inflow orifice diameters results in an underestimation of the effective orifice diameter in the smaller sizes of prosthetic valves. Furthermore, in itself this dichotomy in two sizes of outlet tubes is a regrettable simplification, as ideally each internal diameter should be matched to the appropriate outlet tube.

Regrettable also is the fact that the results were published without warning the reader about their flawed nature and without any explanation why the Working Group departed from their own published stipulations in the ISO PAS (2). Anonymisation of the various individual valvar prostheses could have avoided any misperception about the published, but only partly correct EOA’s.

We strongly suggest that this study should be repeated whilst adhering to the PAS 7020 and a revision of the study should be published. If not, we request the published EOA’s to be retracted, since it consists of flawed results, by which results of a partially flawed test-up only adds to the already confusing information in valve sizing.


Acknowledgments

None.


Footnote

Funding: None.

Conflicts of Interest: The authors have no conflicts of interest to declare.

Open Access Statement: This is an Open Access article distributed in accordance with the Creative Commons Attribution-NonCommercial-NoDerivs 4.0 International License (CC BY-NC-ND 4.0), which permits the non-commercial replication and distribution of the article with the strict proviso that no changes or edits are made and the original work is properly cited (including links to both the formal publication through the relevant DOI and the license). See: https://creativecommons.org/licenses/by-nc-nd/4.0/.


References

  1. Wu C, Green C, Marquez S, et al. Effective orifice diameter: a new sizing parameter of surgical valve prostheses to inform valve selection. Ann Cardiothorac Surg 2024;13:283-90. [Crossref] [PubMed]
  2. International Organization for Standardization. ISO/PAS 7020:2023. Sizing parameters of surgical valve prostheses: Requirements regarding the application of ISO 5840-2. Geneva: ISO; 2023.
Cite this article as: van Boxtel AGM, Ebels T. Testing errors in measuring the effective orifice diameter as a new sizing parameter of surgical valve prostheses. Ann Cardiothorac Surg 2026;15(5):77. doi: 10.21037/acs-2026-1-0038

Article Options

Download Citation